Why Does My Temperature Data Logger Show Different Readings Malaysia?
Why Does My Temperature Data Logger Show Different Readings Malaysia?
You place two temperature meters side by side.
One shows:
25.2°C
The other shows:
26.0°C
Does this mean one of them is faulty?
Not necessarily.
Differences between temperature and humidity instruments can occur because of accuracy specifications, sensor position, response time, airflow, direct heat exposure, stabilisation time and calibration condition.
This is particularly important when comparing a Benetech GM1365 Temperature & Humidity Data Logger with another thermo-hygrometer or environmental monitoring instrument.
Before deciding that your logger is inaccurate, investigate the measurement conditions first.
1. Every Instrument Has an Accuracy Specification
A digital instrument showing decimal places does not mean every displayed digit represents absolute accuracy.
For example:
Display Resolution ≠ Measurement Accuracy
An instrument may display:
25.3°C
but its specified accuracy must still be considered.
When comparing two instruments, check the manufacturer's accuracy specifications for both.
Two correctly operating instruments can show slightly different values and still remain within their respective specifications.
2. Resolution and Accuracy Are Different
This is one of the most misunderstood concepts in measurement.
Resolution
The smallest change shown by the display.
Accuracy
How closely the measurement is expected to represent the actual value under specified conditions.
A meter displaying more decimal places is not automatically more accurate.
Therefore, don't choose a data logger simply because:
“This one shows more digits.”
Look at the actual measurement specification.
3. Sensor Position Can Change the Reading
Even a small difference in position can matter.
Imagine:
Logger A → On a table
Logger B → Near an air-conditioning outlet
The readings may differ because they are not measuring exactly the same environment.
Other problematic locations include:
Near windows
Near external walls
Direct sunlight
Near machinery
Near doors
Beside heaters
Near ventilation outlets
Close to cold surfaces
Before comparing two instruments, place them as close together as practical under stable conditions.
4. Give the Instruments Time to Stabilise
Suppose you move a logger from:
Air-Conditioned Office
to
Warm Warehouse
The sensor will not necessarily display the final environmental condition instantly.
Allow sufficient time for the instrument to stabilise.
Comparing readings immediately after moving instruments between environments can produce misleading conclusions.
5. Different Sensors Have Different Response Times
Two instruments exposed to the same environmental change may respond at different speeds.
For example, when moving from a cool room into a warmer area:
Meter A → Responds quickly
Meter B → Responds more gradually
During the transition, their readings may differ.
After sufficient stabilisation, the difference may become much smaller.
This is why comparison should be performed under stable conditions.
6. Airflow Can Affect Measurement
Air-conditioning and ventilation can create local differences.
A sensor directly exposed to moving air may respond differently from one located behind an object or in a less ventilated position.
This is particularly important when investigating:
HVAC systems
Clean rooms
Laboratories
Warehouses
Equipment rooms
The measurement location should represent the environment you actually want to monitor.
7. Direct Sunlight Can Cause Problems
If a logger is exposed directly to sunlight, the instrument housing itself may absorb heat.
The resulting reading may not represent general ambient air temperature.
This is particularly important in:
Agriculture
Greenhouses
Warehouses near windows
Outdoor-adjacent areas
If the objective is to measure ambient air conditions, sensor placement should minimise inappropriate direct radiant heat exposure.
8. Don't Hold the Sensor While Comparing
Your body generates heat.
Holding a small temperature or humidity instrument can influence the local conditions around the sensor.
For better comparison:
Place Instruments Together → Do Not Hold Them → Allow Stabilisation → Compare
This provides a more meaningful result.
9. Humidity Comparisons Can Be More Difficult
Relative humidity readings can be particularly sensitive to environmental conditions.
Humidity can change with:
Temperature
Airflow
Sensor exposure
Condensation
Movement between rooms
Human proximity
Therefore, seeing different humidity readings immediately after moving two instruments does not automatically mean one sensor is defective.
Allow sufficient stabilisation time.
10. Compare Under Stable Conditions
A simple preliminary comparison can be performed by placing both instruments:
Close together
Away from direct sunlight
Away from HVAC outlets
Away from heat sources
In a relatively stable environment
Then allow them to stabilise before comparing readings.
This is much more meaningful than comparing one instrument in your hand with another instrument mounted on a wall.
11. Is Your Reference Instrument Actually Accurate?
This is another common mistake.
A customer says:
“My GM1365 is wrong because my other meter says 25°C.”
But how do we know the other meter is correct?
If neither instrument has a known calibration status, comparing them only tells you:
The two instruments are different.
It does not establish which one is closer to the reference value.
12. What Is a Better Way to Verify Accuracy?
For measurements that matter, comparison should be made against an appropriate reference under controlled conditions.
Calibration provides a more formal approach.
A calibration process can help determine:
Reference Value
vs
Instrument Indication
and establish the observed measurement error under the calibration conditions.
This is much more meaningful than comparing two unknown meters.
13. When Should You Consider Calibration?
Calibration becomes particularly important when measurements are used for:
QA/QC
Laboratory monitoring
Food storage records
Customer requirements
Environmental documentation
Controlled processes
Audit requirements
The required calibration points and traceability should follow the application and quality system.
14. What If the Logger Was Dropped?
Physical shock can potentially affect an instrument.
If a logger has:
Been dropped
Been exposed to water beyond its intended protection
Experienced extreme conditions
Been stored improperly
Suffered physical damage
and the readings suddenly become suspicious, inspection or calibration may be appropriate.
15. What If the Humidity Sensor Gets Wet?
Condensation or inappropriate moisture exposure can affect humidity measurements.
If a logger is moved rapidly between very different temperature environments, condensation may occur under certain conditions.
Follow the manufacturer's operating and environmental limitations.
If the sensor continues to behave abnormally after returning to appropriate conditions, further inspection may be required.
16. Data Logger Placement Can Create “False Problems”
Imagine a warehouse has three loggers.
Logger A → 27°C
Logger B → 28°C
Logger C → 31°C
The immediate reaction might be:
“Logger C is inaccurate.”
But then you discover Logger C is mounted near the roof under a metal structure receiving substantial heat.
The logger may actually be showing a genuine local hot spot.
This is exactly why data loggers are useful.
Different readings can reveal real environmental differences.
17. Use Multiple Loggers to Find Environmental Differences
Instead of assuming different readings are errors, investigate the pattern.
For example:
Loading Door → Logger A
Centre Warehouse → Logger B
Upper Rack → Logger C
Lower Rack → Logger D
If Logger C repeatedly records higher temperatures during the afternoon, the result may indicate a real environmental pattern rather than instrument failure.
18. GM1365 for Temperature & Humidity Trend Monitoring
The Benetech GM1365 is useful because it does not only provide one current measurement.
It records environmental conditions over time.
This allows users to distinguish between:
Random Reading Difference
and
Repeated Environmental Pattern
A repeated trend can be much more informative than comparing two displays for a few seconds.
Troubleshooting Checklist
If your GM1365 or another temperature data logger appears to show a different reading, check:
1. Are both instruments measuring at the same location?
2. Have both instruments stabilised?
3. Are they exposed to the same airflow?
4. Is either instrument in direct sunlight?
5. Is one near a heat or cooling source?
6. What are the accuracy specifications?
7. What is the calibration status?
8. Is the reference instrument itself trustworthy?
9. Has either instrument been damaged?
10. Does the difference remain consistent over time?
Only after checking these factors should you conclude that the logger may require inspection or calibration.
Different Reading Does Not Automatically Mean Faulty
Remember:
Instrument A ≠ Instrument B
does not automatically mean:
Instrument A = Faulty
or
Instrument B = Faulty
The difference must be evaluated against:
Specification + Environment + Placement + Stabilisation + Calibration
This is the correct measurement approach.
Looking for Temperature & Humidity Logger Calibration in Malaysia?
MTM Precision supplies the Benetech GM1365 Temperature & Humidity Data Logger and a wide range of environmental measuring instruments in Malaysia.
If your logger readings appear inconsistent, send us:
Instrument Model + Application + Reading Difference + Comparison Instrument + Calibration Requirement
We can help determine whether the issue is likely related to application, placement, instrument specification or whether further calibration/inspection should be considered.
Contact MTM Precision
MTM Precision Sdn Bhd
No. 29-1 & 29-2, Jalan Bandar 18,
Pusat Bandar Puchong,
47160 Puchong, Selangor, Malaysia.
Website: www.mtmpre.com.my
WhatsApp: +6016-660 7346
Email: mtmpre@yahoo.com
Different Reading? → Check Placement → Stabilise → Compare Specifications → Verify Calibration — MTM Precision Malaysia
Aug 23,2026