How to Check Auto Level Accuracy with a Two-Peg Test Malaysia
How to Check Auto Level Accuracy with a Two-Peg Test Malaysia
How do you know whether an auto level is still accurate?
One of the most useful field checks is the two-peg test.
The two-peg test is commonly used to check whether an auto level has excessive line-of-sight or collimation error.
It can be particularly useful when:
Readings suddenly become inconsistent
The auto level has been dropped
Benchmark results do not agree
The instrument has been transported frequently
You want to check the equipment before an important project
You suspect the instrument requires calibration
A two-peg test is a useful field check, but it should not be confused with professional calibration.
What Is a Two-Peg Test?
The principle is straightforward.
Two stable points are established some distance apart.
Readings are first taken with the auto level positioned approximately midway between the two points.
The instrument is then moved much closer to one point and another set of readings is taken.
Comparing the two sets of observations helps identify whether the instrument's line of sight introduces excessive error.
Why Position the Auto Level in the Middle First?
When the instrument is positioned approximately halfway between the two staff positions, the sight distances are similar.
This is important because certain line-of-sight errors affect both observations in a similar way.
By comparing the two staff readings, you can establish a more reliable reference height difference between the points.
This becomes the basis for the second part of the test.
Equipment Required
A basic two-peg test requires:
Auto level
Stable survey tripod
Levelling staff
Two stable observation points
Suitable measuring distance
Field notes or recording method
Before beginning, make sure the tripod and levelling staff are in good condition.
A loose tripod or damaged staff can introduce additional errors and make the test less useful.
Step 1: Establish Two Stable Points
Select two stable points on reasonably level ground.
The exact test distance may depend on your procedure, instrument and site conditions.
The important point is to create sufficient separation for the test while maintaining clear staff visibility.
Label the points:
Point A and Point B
Step 2: Set the Auto Level Approximately Midway
Position the tripod approximately halfway between Point A and Point B.
Mount the auto level securely.
Centre the circular bubble and allow the compensator to settle.
Make sure:
Tripod is stable
Instrument is securely mounted
Staff is vertical
Image is properly focused
Parallax is minimised
Step 3: Take the First Two Readings
Place the levelling staff on Point A and record the reading.
Then place it on Point B and record the second reading.
For example:
Reading A1
Reading B1
The difference between these readings provides the reference height difference for the test.
Record everything carefully.
A recording mistake can make the entire test meaningless.
Step 4: Move the Auto Level Closer to One Point
Now relocate the instrument much closer to one of the two points, for example Point A.
Set up and level the instrument again.
Do not move the original Point A and Point B locations.
Step 5: Take the Second Set of Readings
Take another reading on Point A.
Then observe Point B.
Record these as:
Reading A2
Reading B2
Because the sight distances are now significantly unequal, line-of-sight error becomes easier to detect when the results are compared with the reference height difference.
Step 6: Compare the Results
Compare the height difference obtained from the first setup with the result from the second setup.
If the difference exceeds the acceptable tolerance for your instrument or work requirement, the auto level may require adjustment.
Do not use an arbitrary tolerance.
Refer to:
Manufacturer specifications
Company quality procedures
Project requirements
Applicable calibration requirements
Different auto levels and applications may have different acceptable limits.
What If the Auto Level Fails the Two-Peg Test?
Do not immediately assume the instrument is beyond repair.
A failed test may indicate that the instrument requires:
Inspection → Adjustment → Calibration
If there is internal damage, the process may instead be:
Inspection → Repair → Adjustment → Calibration → Verification
The exact solution depends on the instrument's condition.
Should I Adjust the Auto Level Myself?
Some auto levels provide adjustment points for trained users or technicians.
However, randomly turning adjustment screws is not recommended.
Incorrect adjustment can:
Increase the error
Affect instrument geometry
Make troubleshooting more difficult
Create additional calibration problems
If you are uncertain about the correct procedure, have the instrument checked professionally.
Two-Peg Test After Dropping an Auto Level
If an auto level has been dropped, a two-peg test can provide a useful initial indication of whether accuracy has changed.
However, impact can affect more than collimation.
A drop may also damage or disturb:
Compensator
Circular bubble
Focusing mechanism
Optical components
Mechanical components
Therefore, passing one field test does not necessarily prove that every part of the instrument is in perfect condition.
For important work, professional inspection may still be appropriate.
Two-Peg Test vs Calibration
These two terms should not be treated as interchangeable.
Two-Peg Test
A field procedure used to check for possible line-of-sight error.
Professional Calibration
A controlled process used to inspect and verify instrument performance according to the applicable service or calibration procedure.
If your company or project requires documented calibration, a field two-peg test alone may not satisfy that requirement.
How Often Should You Perform a Field Check?
The appropriate frequency depends on how the auto level is used.
Consider additional checking:
Before critical projects
After accidental impact
After rough transportation
When readings appear suspicious
After extended heavy use
Between scheduled calibration intervals
Regular field checks can help identify problems before they create expensive construction errors.
Don't Forget the Tripod and Staff
If a two-peg test produces strange results, also inspect the accessories.
Check the tripod for:
Loose legs
Worn clamps
Unstable head
Movement during observations
Check the levelling staff for:
Bent sections
Loose joints
Damaged locks
Difficulty remaining vertical
Surveying accuracy is produced by the complete measurement system.
Auto Level Calibration & Repair Malaysia
If your auto level fails an accuracy check, MTM Precision can inspect the instrument and advise on the appropriate next step.
Depending on its condition, this may involve:
Inspection
Adjustment
Calibration
Repair
Replacement
Upgrade
MTM Precision also supplies professional auto levels, tripods, levelling staffs and other surveying accessories.
Protect Your Surveying Accuracy
A two-peg test is a valuable part of preventive equipment management.
Instead of waiting until a construction level is obviously wrong, establish a routine:
Use → Field Check → Identify Error → Calibrate → Repair if Necessary → Verify → Return to Site
This helps contractors, engineers and surveyors reduce the risk of inaccurate levelling work.